断续流动原子荧光法测定铅锡合金中的痕量锑
Determination of Trace Amount of Antimony in Lead-tin Alloy by Intermittent Flow Atomic Fluorescence Spectrometry

作  者: ; ; ;

机构地区: 惠州学院化学工程系

出  处: 《广东化工》 2012年第2期187-188,共2页

摘  要: 目的:建立氢化物发生双道原子荧光法测定铅锡合金中锑含量的方法,为测定铅锡合金中锑提供依据。方法:选择了最佳的仪器条件,进行样品处理方法、KBH4浓度、检出限、回收率等研究。结果:本方法的标准曲线方程y=157.56x-2.3962,相关系数γ=0.9999,线性范围为0~40μg/L,精密度为1.3%,检出限为0.0304μg/L,回收率达115%。结论:本方法操作简单、快速、灵敏度高、干扰少,可获得满意的结果。 Purpose: A analytical method is recommended for determination of Trace Amount of Antimony in Lead-Tin Alloy by hydride generation double channel atomic fluorescence spectrometry,which provides basis for its measurement.Methods: Optimum condition for the instrument determination is applied to conduct researches relating to the method of sampling,concentration of KBH4,detection limits and recovery rate.Results: The standard curve of the linear equation was y=157.56x-2.3962 with correlation coefficient γ = 0.9999 and linear range of 0~40 μg/L.The precision and the detection limits were 1.3 % and 0.0304 μg/L respectively while the returns-ratio achieves 115 %.Conclusion: The method is easy,fast,highly sensitive and with no much interference during the operation,which can lead to satisfactory results.

关 键 词: 原子荧光法 铅锡合金 研究