摘 要: 研究了钒 荧光镓(LMG)络合物在碳糊电极上的吸附伏安行为,发现其在+0.81V(vs.SCE)的阳极二次导数峰电流与钒(V)的浓度在一定范围内呈线性关系。在pH=4.3的0.26mol/L的HAc NH4Ac缓冲溶液中,于+0.4V富集,从+0.4至+1.4V以300mV/s线性扫描,方法线性范围为4.0×10-9~1.0×10-7mol/L(5.0×10-7mol/LLMG)和1.0×10-7~1.2×10-6mol/L(2.0×10-6mol/LLMG),检出限为1.0×10-9mol/L(S/N=3)。该法用于煤飞灰,粉煤灰和水样中钒的测定,结果满意。 A new method was presented for the determination of vanadium based on anodic adsorptive stripping of the vanadium (Ⅴ)-lumogallion (LMG) complex at a carbon paste electrode (CPE). The complex can be adsorbed on the surface of the CPE and an anodic peak at 0.81 V (vs.SCE) can be observed with linear-scanning from 0.4 V to 1.4 V. Optimal analytical conditions were found to be :(0.26) mol/L HAc-NH-4Ac (pH=4.3), 120 s for accumulation at 0.4 V and a 300 mV/s scan rate. The linear range is 4.0×10+(-9)~1.0×10+(-7)mol/L(5.0×10+(-7)mol/L LMG) and 1.0×10+(-7)~1.2×10+(-6)mol/L(2.0×10+(-6)mol/L LMG). The detection limit is as low as 1.0×10+(-9)mol/L (S/N=3). The proposed method has been applied successfully to the determination of vanadium in flying coal-ash, fly-ash and tap water.